Cine SSFP pulse sequence (parallel imaging as required)
2CH, 4CH, SA and LVOT (2 orthogonal) cine images
SA cine stack (from mitral valve to apex)
Slice thickness 6-10mm
Inter-slice gap 0-4mm to equal 10mm
Temporal resolution ≤ 45ms
Tips and Tricks
To reduce breath-hold times use acceleration techniques
Contouring
In a healthy heart there is usually one less slice to contour in end- systole at the base of the heart (longitudinal LV shortening). Correlate SA to long axis view if available to identify mitral valve plane.
Use the movie function of the analysis software for correct alignments
Different methods have been proposed to deal with trabeculation and papillary muscles. Use a consistent approach and the correct normal values for the chosen method.