Description
  1. Cine SSFP pulse sequence (parallel imaging as required)
  2. 2CH, 4CH, SA and LVOT (2 orthogonal) cine images
  3. SA cine stack (from mitral valve to apex)

    Slice thickness 6-10mm

    Inter-slice gap 0-4mm to equal 10mm

  4. Temporal resolution ≤ 45ms
Tips and Tricks
  1. To reduce breath-hold times use acceleration techniques
  2. Contouring
    • In a healthy heart there is usually one less slice to contour in end- systole at the base of the heart (longitudinal LV shortening). Correlate SA to long axis view if available to identify mitral valve plane.
    • Use the movie function of the analysis software for correct alignments
    • Different methods have been proposed to deal with trabeculation and papillary muscles. Use a consistent approach and the correct normal values for the chosen method.