- Segmented IR GRE imaging during diastolic rest period
- 4CH, 3CH, 2CH, SA images
- In-plane resolution : <2mm
- EGE: image 1-3min after contrast, TI >400ms
- LGE: ≥10min after Gd injection (0.1 – 0.2mmol /kg))
- The delay may be shorter if lower Gd doses are used
- The delay may be increased in a low output state
- TI set to null normal myocardium:
- TI scout or Look Locker sequence
- Phase-sensitive sequence with fixed TI as alternative
- Readout:
- Usually every other heartbeat
- Every heartbeat in the setting of bradycardia
- Every third heartbeat in the setting of tachycardia
|